Cardiac Antibody | TechNote

CARDIAC HIGH-THROUGHPUT ASSAY TECHNOLOGIES

Comparison of fluorescent and chemiluminescent immunoassay technologies used in high-throughput cardiac biomarker testing

FEATURE

FIA | Fluorescence Immunoassay

CLIA | Chemiluminescent Immunoassay Enzyme-labeled antibodies catalyze light-emitting reactions measured without external excitation

Fluorophore-labeled antibodies emit light upon optical excitation; signal intensity proportional to analyte concentration

DETECTION PRINCIPLE

Fluorescence emission

Chemiluminescence emission

SIGNAL TYPE

Typically picogram per milliliter (pg/mL) range; may be limited by background fluorescence or optical noise Determined by antibody affinity & assay design; generally comparable to CLIA Used in certain high-sensitivity & electro- chemiluminescent (ECL) platforms; including Roche Cobas analyzers

Often femtogram per milliliter (fg/ mL) range; high signal amplification & broader dynamic range

DETECTION SENSITIVITY

Also antibody-dependent; comparable specificity to FIA

SPECIFICITY

Standard in automated high-throughput systems such as Abbott Architect, Siemens ADVIA Centaur, Roche Cobas e-Series, Beckman Coulter Access Very high sensitivity, wide dynamic range, optimized for continuous, high-volume testing

COMMON APPLICATIONS

High sensitivity, rapid detection, suitable for compact analyzers

PERFORMANCE SUMMARY

EVALUATION METHOD

Antibody pairs specific to cTnI , CRP , NT-proBNP , and CK-MB were evaluated in a two-step sandwich format using either a fluorescence immunoassay (FIA) or a chemiluminescent immunoassay (CLIA). In FIA, capture antibodies were coated on a nitrocellulose (NC) membrane and detection antibodies were conjugated to Europium. In CLIA, capture antibodies were coated on magnetic beads and detection antibodies were conjugated to alkaline phosphatase (ALP). The limit of detection (LOD) was determined for each antibody pair.

TWO-STEP SANDWICH ASSAY

DETECTION ANTIBODY CONJUGATE FIA | Ab-Europium Conjugate CLIA | Ab-AKP Conjugate

ADD SAMPLE

ADD DETECTOR

ANTIGEN

FIA | Coated on NC Membrane CLIA | Coated on Magnetic Bead CAPTURE ANTIBODY

Powered by