Meridian Lyo-Ready Direct LAMP Saliva Mixes

High inhibitor tolerance and fast amplification to deliver quicker time-to-results (TTR) in saliva, sputum and samples containing Universal Transport Media (UTM)

Lyo-Ready ™ Direct DNA LAMP Saliva (MDX134)

CMV

ADV

70

70

60

60

50

50

40

40

30

30

20

20

10

10

0

0

0% inhibitor

30% saliva

30% UTM + Sputum

30% VTM

NTC

0% inhibitor

30% saliva

30% UTM + Sputum

30% VTM

NTC

HSV2

Primers for CMV, ADV or HSV-2 were added to Lyo-Ready ™ Direct DNA LAMP Saliva (MDX134) and lyophilized ( blue ) or added to NEB WarmStart ® ( orange ) and kept as a liquid. The mixes were amplified using inactivated Cytomegalovirus (CMV), Adenovirus (ADV) or Herpes simplex virus 2 (HSV2). Reactions were prepared in the presence of 30% human saliva, 30% Universal Transport Medium (UTM) + an artificial sputum swab and Viral Transport Medium (VTM). All reactions were incubated at 65 °C for 60 minutes, and average time to results (TTR) was measured at 1:10 of end fluorescence. Error bars represent the standard deviation across four technical replicates. Results show that MDX134 ( blue ) has a shorter TTR than NEB WarmStart ® ( orange ) amplification across all three sample types. The data suggests that MDX134 performs better with faster amplification across the range of inhibitors.

70

60

50

40

30

20

10

0

0% inhibitor

30% saliva

30% UTM + Sputum

30% VTM

NTC

Lyophilized | NEB WarmStart ®

Lyo-Ready ™ Direct RNA/DNA LAMP Saliva (MDX135)

RSV

Primers for RSV were added to Lyo-Ready ™ Direct RNA/DNA LAMP Saliva (MDX135) and lyophilized ( blue ) or added to NEB WarmStart ® ( orange ) and kept as a liquid. The mixes were amplified using inactivated respiratory syncytial virus (RSV) RNA prepared with 30% human saliva, 30% Universal Transport Medium (UTM) with/without an artificial sputum swab and Viral Transport Medium (VTM). All reactions were incubated at 65 °C for 60 minutes, and the average time to results (TTR) was measured at 1:10 of end fluorescence. Error bars represent the standard deviation across six technical replicates. Results show that MDX135 ( blue ) has a shorter TTR than NEB WarmStart ® ( orange ) amplification across all sample types. The data suggests that MDX135 performs better with faster amplification across the range of inhibitors.

70

60

50

40

30

20

10

Lyophilized | NEB WarmStart ®

0

0% inhibitor

30% Saliva

30% UTM

30% VTM

30% Sputum in UTM

30% Sputum

NTC

Powered by